Under Pressure: Making Your Autoclave Work Smarter
Updated: 5 days ago
In a previous Greening the Bench post, Every Drop Counted: Why Water Metering Matters in the Lab, we looked at how to find out where your lab's water actually goes. The autoclave was among the usual suspects at the top of the list.
Autoclaves are essential. They sterilise media, glassware and equipment, and they make biological waste safe before it leaves the building. That job is non-negotiable, and nothing in this blog is about doing it less thoroughly. However, what we can do is make sure every cycle you run is one that needed running, and that the machine isn't using energy and water when it isn't doing anything at all.
Why the Autoclave Deserves Your Attention
Few pieces of lab equipment use energy and water on the scale that an autoclave does. Every cycle heats a pressurised chamber, typically to 121°C or 134°C, and in many models cold mains water is then used to cool the steam and condensate before it can safely go down the drain.
That combination is what makes the autoclave stand out. A ULT freezer uses a lot of energy and single-pass cooling uses a lot of water, but an autoclave can do both, often several times a day. If you carried out a water audit after our metering post, there's a good chance the autoclave was near the top of your list.
The upside is that small changes in how you use it multiply across every cycle, every week, all year, which can lead to lower utility bills, less wear on the machine and fewer maintenance call-outs, alongside the carbon and water savings.
Fill It Up with Full Loads
Just like a washing machine you'll find at home, an autoclave spends a large share of its energy and water on each cycle whether the chamber is full or half empty. Two half loads cost roughly twice as much as one full one, so batching is the quickest win available.
A few ways to get more out of every cycle:
🗓️ Agree set run times: Schedule cycles that suit your lab, for example mid-morning and late afternoon, so people can batch their items rather than running the autoclave whenever something is ready.
🧺 Share the chamber: A shared holding area lets several users or groups combine loads, as long as the materials are compatible with the same cycle.
📋 Keep a simple log: Recording what went into each cycle quickly shows how often the machine runs part full.
That being said, it is important that the chamber isn't 'overfilled'. Steam needs to reach every surface, so always follow the loading limits and patterns in your procedures and the manufacturer's guidance. An overloaded chamber that fails it's sterilisation check costs far more than it saves. If you plan to change how you load or combine items, check with your biosafety officer first, as validated cycles are tied to specific load types and sizes.
The Right Cycle for the Job
Most autoclaves offer several programmes, and they aren't interchangeable. Using the longest, hottest cycle for everything "just to be safe" wastes energy and time, and can damage heat-sensitive items.
Work with your biosafety officer or lab manager to make sure each material type has a validated cycle, and that it's clearly labelled on or beside the machine.
Does It Really Need Sterilising?
Often, not everything that goes into the autoclave needs to be there. Common examples include clean glassware heading for non-sterile use and non-hazardous waste bagged for autoclaving out of habit.
Anything your risk assessment, local rules or waste regulations say must be autoclaved must be autoclaved, every time. But a quick review of what goes through the machine, checked against those requirements, can often uncover items that are there by default rather than necessity. Removing them frees up capacity, so the loads that do need the autoclave get through faster.
Mind the Idle Time
Many autoclaves, especially jacketed models, keep the jacket or steam generator hot between cycles so they're ready to go. That's convenient during a busy day, but overnight and at weekends it's energy spent keeping a hot machine waiting for nobody.
A few things worth checking:
⏻ Standby and timers: See whether your model has an eco or standby mode, or a programmable start so it warms up just before the first scheduled run.
🌙 End-of-day shutdown: If it's safe and your facilities team agrees, add the autoclave to your lab's shutdown checklist. Many autoclaves need an empty warm-up cycle after a cold start, which uses some energy and water, so switching off pays off best overnight and at weekends rather than for short gaps during the day.
🔧 Maintenance: Worn door seals, faulty steam traps and leaking valves waste both steam and water, so keep servicing up to date.
Cooling Water: The Hidden Tap
Steam and condensate leaving an autoclave are too hot to discharge directly, so many units cool them by mixing in cold mains water, known as drain cooling (sometimes called drain tempering). In some installations this water runs continuously, whether a cycle is running or not, and it can become one of the largest water uses in the lab.
Water saving options exist for both new and existing machines. Condensate cooling systems use a heat exchanger or a closed-loop recirculating chiller to cool the discharge, so only a fraction of the mains water is needed, and some suppliers offer retrofit kits for older units. If a system uses the building's chilled water supply, bear in mind it shifts some of the work onto the building's chiller, so part of the saving comes back as energy use, though it's usually still a worthwhile trade. Others add a valve that only lets cooling water flow when the drain is actually hot. If you're replacing an autoclave, ask suppliers about water use per cycle and at idle, not just chamber size and cycle time.
These are usually conversations for your facilities team, and that's where your metering data becomes powerful. A measured figure in litres per day makes a far stronger case than an example online, and the payback on a cooling retrofit can be calculated, and is often quicker than people expect.
Action of the Week
Run a one-week autoclave log. Put a sheet next to the machine and ask everyone to record each cycle: the date, the programme used, roughly how full the chamber was, and what was in it.
At the end of the week, count how many cycles ran less than half full. That one number tells you how much room there is for batching, and gives you a clear starting point for agreeing shared run times with your team.
Conclusion: Smarter Cycles, Same Safety
The autoclave is one of the hardest-working machines in the lab, and rightly so: sterility and safe decontamination always come first. But within your validated protocols there's plenty of room to work smarter: fuller loads, the right cycle, only what genuinely needs sterilising, less idle time and less water down the drain. Each change saves energy, water and money, and keeps the machine free for the cycles that matter.
Frequently Asked Questions
Why is the autoclave one of the biggest energy and water users in a lab?
Every cycle generates high-temperature, pressurised steam, which takes a significant amount of energy, and many models also use cold mains water to cool steam and condensate before it can go to drain. Because it uses both energy and water, often several times a day and sometimes even while idle, the autoclave is frequently one of the largest combined consumers in a lab.
Is it safe to run fuller autoclave loads?
Yes, provided you stay within the loading limits and patterns set out in your validated procedures and the manufacturer's guidance. Steam has to reach every surface, so full should never mean overpacked. Batching compatible items into fewer, well-loaded cycles saves energy and water without compromising sterilisation.
How can I reduce my autoclave's water consumption?
Start by finding out whether your autoclave runs drain cooling water continuously. Condensate cooling systems that use a heat exchanger or recirculating chiller, retrofitted water-saving kits and temperature-controlled drain cooling valves can all substantially reduce mains water use. Fixing faulty steam traps and seals, and running fewer part loads, helps too.
Should autoclaves be switched off overnight?
Many autoclaves, especially jacketed models, keep heating between cycles so they are ready to run. Where it is safe and agreed with your facilities team, using a standby mode, a programmable start time or an end-of-day shutdown can cut idle energy use. This works best for longer gaps such as overnight and weekends, as many units need a warm-up cycle after a cold start. Always check with facilities and the manufacturer's guidance first.